Browsing by Author "Bhonsle, Suyashree P."
Now showing 1 - 4 of 4
Results Per Page
Sort Options
- Ablation outcome of irreversible electroporation on potato monitored by impedance spectrum under multi-electrode systemZhao, Yajun; Liu, Hongmei; Bhonsle, Suyashree P.; Wang, Yilin; Davalos, Rafael V.; Yao, Chenguo (2018-09-20)Background Irreversible electroporation (IRE) therapy relies on pulsed electric fields to non-thermally ablate cancerous tissue. Methods for evaluating IRE ablation in situ are critical to assessing treatment outcome. Analyzing changes in tissue impedance caused by electroporation has been proposed as a method for quantifying IRE ablation. In this paper, we assess the hypothesis that irreversible electroporation ablation outcome can be monitored using the impedance change measured by the electrode pairs not in use, getting more information about the ablation size in different directions. Methods Using a square four-electrode configuration, the two diagonal electrodes were used to electroporate potato tissue. Next, the impedance changes, before and after treatment, were measured from different electrode pairs and the impedance information was extracted by fitting the data to an equivalent circuit model. Finally, we correlated the change of impedance from various electrode pairs to the ablation geometry through the use of fitted functions; then these functions were used to predict the ablation size and compared to the numerical simulation results. Results The change in impedance from the electrodes used to apply pulses is larger and has higher deviation than the other electrode pairs. The ablation size and the change in resistance in the circuit model correlate with various linear functions. The coefficients of determination for the three functions are 0.8121, 0.8188 and 0.8691, respectively, showing satisfactory agreement. The functions can well predict the ablation size under different pulse numbers, and in some directions it did even better than the numerical simulation method, which used different electric field thresholds for different pulse numbers. Conclusions The relative change in tissue impedance measured from the non-energized electrodes can be used to assess ablation size during treatment with IRE according to linear functions.
- Effects of internal electrode cooling on irreversible electroporation using a perfused organ modelO’Brien, T. J.; Bonakdar, Mohammad; Bhonsle, Suyashree P.; Neal, Robert E.; Aardema, C.H.; Robertson, John L.; Goldberg, S.N.; Davalos, Rafael V. (Taylor and Francis Ltd, 2018-05-28)Purpose: This study evaluates the effects of active electrode cooling, via internal fluid circulation, on the irreversible electroporation (IRE) lesion, deployed electric current and temperature changes using a perfused porcine liver model. Materials and methods: A bipolar electrode delivered IRE electric pulses with or without activation of internal cooling to nine porcine mechanically perfused livers. Pulse schemes included a constant voltage, and a preconditioned delivery combined with an arc-mitigation algorithm. After treatment, organs were dissected, and treatment zones were stained using triphenyl-tetrazolium chloride (TTC) to demonstrate viability. Results: Thirty-nine treatments were performed with an internally cooled applicator and 21 with a non-cooled applicator. For the constant voltage scenario, the average final electrical current measured was 26.37 and 29.20 A for the cooled and uncooled electrodes respectively (p≤.001). The average final temperature measured was 33.01 and 42.43 °C for the cooled and uncooled electrodes respectively (p≤.0001). The average measured ablations (fixed lesion) were 3.88-by-2.08 cm and 3.86-by-2.12 cm for the cooled and uncooled electrode respectively (p≤.2495, p≤.7507). Similarly, the preconditioned/arc-mitigation scenario yielded an average final electrical current measurement of a 41.07 and 47.20 A for the cooled and uncooled electrodes respectively (p≤.0001). The average final temperature measured was 34.93 and 44.90 °C for the cooled and uncooled electrodes respectively (p≤.0001). The average measured ablations (fixed lesion) were 3.67-by-2.27 cm and 3.58-by-2.09 cm for the cooled and uncooled applicators ((p≤.7906; p≤.5595)). Conclusions: The internally-cooled bipolar applicator offers advantages that could improve clinical outcomes. Thermally mitigating internal perfusion technology reduced tissue temperatures and electric current while maintaining similar lesion sizes.
- Mitigation of impedance changes due to electroporation therapy using bursts of high-frequency bipolar pulsesBhonsle, Suyashree P.; Arena, Christopher B.; Sweeney, Daniel C.; Davalos, Rafael V. (2015-08-27)Background For electroporation-based therapies, accurate modeling of the electric field distribution within the target tissue is important for predicting the treatment volume. In response to conventional, unipolar pulses, the electrical impedance of a tissue varies as a function of the local electric field, leading to a redistribution of the field. These dynamic impedance changes, which depend on the tissue type and the applied electric field, need to be quantified a priori, making mathematical modeling complicated. Here, it is shown that the impedance changes during high-frequency, bipolar electroporation therapy are reduced, and the electric field distribution can be approximated using the analytical solution to Laplace's equation that is valid for a homogeneous medium of constant conductivity. Methods Two methods were used to examine the agreement between the analytical solution to Laplace's equation and the electric fields generated by 100 µs unipolar pulses and bursts of 1 µs bipolar pulses. First, pulses were applied to potato tuber tissue while an infrared camera was used to monitor the temperature distribution in real-time as a corollary to the electric field distribution. The analytical solution was overlaid on the thermal images for a qualitative assessment of the electric fields. Second, potato ablations were performed and the lesion size was measured along the x- and y-axes. These values were compared to the analytical solution to quantify its ability to predict treatment outcomes. To analyze the dynamic impedance changes due to electroporation at different frequencies, electrical impedance measurements (1 Hz to 1 MHz) were made before and after the treatment of potato tissue. Results For high-frequency bipolar burst treatment, the thermal images closely mirrored the constant electric field contours. The potato tissue lesions differed from the analytical solution by 39.7 ± 1.3 % (x-axis) and 6.87 ± 6.26 % (y-axis) for conventional unipolar pulses, and 15.46 ± 1.37 % (x-axis) and 3.63 ± 5.9 % (y-axis) for high- frequency bipolar pulses. Conclusions The electric field distributions due to high-frequency, bipolar electroporation pulses can be closely approximated with the homogeneous analytical solution. This paves way for modeling fields without prior characterization of non-linear tissue properties, and thereby simplifying electroporation procedures.
- Non-linearity and Dispersion Effects in Tissue Impedance during Application of High Frequency Electroporation-Inducing Pulsed Electric FieldsBhonsle, Suyashree P. (Virginia Tech, 2018-01-27)Since its conception in 2005, irreversible electroporation (IRE), a non-thermal tumor ablation modality, was investigated for safety and efficacy in clinical applications concerning different organs. IRE utilizes high voltage (~3kV), short duration (~100us) pulses to create transient nanoscale defects in the plasma membrane to cause cell death due to irreversible defects, osmotic imbalances and ATP loss. More recently, high-frequency irreversible electroporation (H-FIRE), which employs narrow bipolar pulses (~0.5-10us) delivered in bursts (on time ~100us), was invented to provide benefits such as the mitigation of intense muscle contractions associated with IRE-based treatments. Furthermore, H-FIRE exhibits the potential to improve lesion predictability in homogeneous and heterogeneous tissue masses. Therapeutic IRE and H-FIRE utilize source and sink electrodes inserted into or around the tumor to deliver the treatment. Prediction of the ablation size, for a set of parameters, can be achieved by the use of pre-treatment planning algorithms that calculate the induced electric field distribution in the target tissue. An electric field above a certain threshold induces cell death and parameters are tuned to ensure complete tumor coverage while sparing the nearby healthy tissue. IRE studies have shown that the underlying field is influenced by the increase in tissue conductivity due to enhanced membrane permeability, and treatment outcome can be improved when this nonlinearity is accounted for in numerical models. Since IRE pulses far exceed the time constant of the cell (~1us), the tissue response can be treated as essentially DC a static approximation can be used to predict the field distribution. Alternately, as H-FIRE pulses are on the order of the time constant of the membrane, the tissue response can no longer be treated as DC. The complexity of the H-FIRE-induced field distribution is further enhanced due to the dispersion and non-linearity in biological tissue impedance during treatment. In this dissertation, we have studied the electromagnetic fields induced in tissue during H-FIRE using several experimental and modeling techniques. In addition, we have characterized the nonlinearity and dispersion in tissue impedance during H-FIRE treatments and proposed simpler methods to predict the field distribution to enable easier translation to the clinic.