Consensus Head Acceleration Measurement Practices (CHAMP): Laboratory Validation of Wearable Head Kinematic Devices

dc.contributor.authorGabler, Leeen
dc.contributor.authorPatton, Declanen
dc.contributor.authorBegonia, Mark T.en
dc.contributor.authorDaniel, Rayen
dc.contributor.authorRezaei, Ahmaden
dc.contributor.authorHuber, Colinen
dc.contributor.authorSiegmund, Gunteren
dc.contributor.authorRooks, Tyleren
dc.contributor.authorWu, Lyndiaen
dc.date.accessioned2022-10-05T17:04:58Zen
dc.date.available2022-10-05T17:04:58Zen
dc.date.issued2022-09en
dc.description.abstractWearable devices are increasingly used to measure real-world head impacts and study brain injury mechanisms. These devices must undergo validation testing to ensure they provide reliable and accurate information for head impact sensing, and controlled laboratory testing should be the first step of validation. Past validation studies have applied varying methodologies, and some devices have been deployed for on-field use without validation. This paper presents best practices recommendations for validating wearable head kinematic devices in the laboratory, with the goal of standardizing validation test methods and data reporting. Key considerations, recommended approaches, and specific considerations were developed for four main aspects of laboratory validation, including surrogate selection, test conditions, data collection, and data analysis. Recommendations were generated by a group with expertise in head kinematic sensing and laboratory validation methods and reviewed by a larger group to achieve consensus on best practices. We recommend that these best practices are followed by manufacturers, users, and reviewers to conduct and/or review laboratory validation of wearable devices, which is a minimum initial step prior to on-field validation and deployment. We anticipate that the best practices recommendations will lead to more rigorous validation of wearable head kinematic devices and higher accuracy in head impact data, which can subsequently advance brain injury research and management.en
dc.description.versionPublished versionen
dc.format.mimetypeapplication/pdfen
dc.identifier.doihttps://doi.org/10.1007/s10439-022-03066-0en
dc.identifier.eissn1573-9686en
dc.identifier.issn0090-6964en
dc.identifier.pmid36104642en
dc.identifier.urihttp://hdl.handle.net/10919/112082en
dc.language.isoenen
dc.publisherSpringeren
dc.rightsCreative Commons Attribution 4.0 Internationalen
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/en
dc.subjectAccuracyen
dc.subjectBest practicesen
dc.subjectHead impact kinematicsen
dc.subjectRecommendationsen
dc.subjectValidationen
dc.titleConsensus Head Acceleration Measurement Practices (CHAMP): Laboratory Validation of Wearable Head Kinematic Devicesen
dc.title.serialAnnals of Biomedical Engineeringen
dc.typeArticle - Refereeden
dc.type.dcmitypeTexten

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
s10439-022-03066-0.pdf
Size:
1.12 MB
Format:
Adobe Portable Document Format
Description:
Published version