Development of a Phantom Tissue for Blood Perfusion Measurements and Noninvasive Blood Perfusion Estimation in Living Tissue

dc.contributor.authorMudaliar, Ashvinikumaren
dc.contributor.committeechairDiller, Thomas E.en
dc.contributor.committeecochairScott, Elaine P.en
dc.contributor.committeememberLanz, Otto I.en
dc.contributor.committeememberTafti, Danesh K.en
dc.contributor.committeememberVick, Brian L.en
dc.contributor.departmentMechanical Engineeringen
dc.date.accessioned2014-03-14T20:08:13Zen
dc.date.adate2007-04-17en
dc.date.available2014-03-14T20:08:13Zen
dc.date.issued2007-01-23en
dc.date.rdate2007-04-17en
dc.date.sdate2007-03-15en
dc.description.abstractA convenient method for testing and calibrating surface perfusion sensors has been developed. A phantom tissue model is used to mimic the non-directional blood flow of tissue perfusion. A computational fluid dynamics (CFD) model was constructed in Fluent to design the phantom tissue and validate the experimental results. The phantom perfusion system was used with a perfusion sensor based on the clearance of thermal energy. A heat flux gage measures the heat flux response of tissue when a thermal event (convective cooling) is applied. The blood perfusion and contact resistance are estimated by a parameter estimation code. From the experimental and analytical results, it was concluded that the probe displayed good measurement repeatability and sensitivity. The experimental perfusion measurements in the tissue were in good agreement with those of the CFD models and demonstrated the value of phantom tissue system. This simple, cost effective, and noninvasive convective blood perfusion system was then tested in animal models. The perfusion system was evaluated for repeatability and sensitivity using isolated rat liver and exposed rat kidney tests. Perfusion in the isolated liver tests was varied by controlling the flow of the perfusate into the liver, and the perfusion in the exposed kidney tests was varied by temporarily occluding blood flow through the renal artery and vein. The perfusion estimated by the convective perfusion probe was in good agreement with that of the metered flow of perfusate into the liver model. The liver tests indicated that the probe can be used to detect small changes in perfusion (0.005 ml/ml/s). The probe qualitatively tracked the changes in the perfusion in kidney model due to occlusion of the renal artery and vein.en
dc.description.degreePh. D.en
dc.identifier.otheretd-03152007-144329en
dc.identifier.sourceurlhttp://scholar.lib.vt.edu/theses/available/etd-03152007-144329/en
dc.identifier.urihttp://hdl.handle.net/10919/26435en
dc.publisherVirginia Techen
dc.relation.haspartdissertation.pdfen
dc.rightsIn Copyrighten
dc.rights.urihttp://rightsstatements.org/vocab/InC/1.0/en
dc.subjectParameter Estimationen
dc.subjectHeat fluxen
dc.subjectContact Resistanceen
dc.subjectPhantom Tissueen
dc.subjectBlood Perfusionen
dc.titleDevelopment of a Phantom Tissue for Blood Perfusion Measurements and Noninvasive Blood Perfusion Estimation in Living Tissueen
dc.typeDissertationen
thesis.degree.disciplineMechanical Engineeringen
thesis.degree.grantorVirginia Polytechnic Institute and State Universityen
thesis.degree.leveldoctoralen
thesis.degree.namePh. D.en