dc.identifier.citationRenner, K.E.; Williams, D.B.; Queen, R.M. The Reliability and Validity of the Loadsol® under Various Walking and Running Conditions. Sensors 2019, 19, 265.
dc.identifier.urihttp://hdl.handle.net/10919/86870
dc.description.abstractThe assessment of loading during walking and running has historically been limited to data collection in laboratory settings or with devices that require a computer connection. This study aims to determine if the loadsol®—a single sensor wireless insole—is a valid and reliable method of assessing force. Thirty (17 male and 13 female) recreationally active individuals were recruited for a two visit study where they walked (1.3 m/s) and ran (3.0 and 3.5 m/s) at a 0%, 10% incline, and 10% decline, with the visits approximately one week apart. Ground reaction force data was collected on an instrumented treadmill (1440 Hz) and with the loadsol® (100 Hz). Ten individuals completed the day 1 protocol with a newer 200 Hz loadsol®. Intraclass correlation coefficients (ICC3,k) were used to assess validity and reliability and Bland–Altman plots were generated to better understand loadsol® validity. Across conditions, the peak force ICCs ranged from 0.78 to 0.97, which increased to 0.84–0.99 with the 200 Hz insoles. Similarly, the loading rate ICCs improved from 0.61 to 0.97 to 0.80–0.96 and impulse improved from 0.61 to 0.97 to 0.90–0.97. The 200 Hz insoles may be needed for loading rate and impulse in running. For both walking and running, the loadsol® has excellent between-day reliability (>0.76).en_US
dc.format.mimetypeapplication/pdf
dc.language.isoen_US
dc.publisherMDPI
dc.rightsCreative Commons Attribution 4.0 International (CC BY 4.0)*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.titleThe Reliability and Validity of the Loadsol® under Various Walking and Running Conditionsen_US
dc.typeArticle - Refereed
dc.date.updated2019-01-24T09:23:02Z
dc.title.serialSensors
dc.identifier.doihttps://doi.org/10.3390/s19020265
dc.type.dcmitypeText


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record

Creative Commons Attribution 4.0 International (CC BY 4.0)
License: Creative Commons Attribution 4.0 International (CC BY 4.0)