The design, development and in vitro testing of an electric motor driven artificial heart

dc.contributor.authorJanney, Joseph T.en
dc.contributor.departmentMechanical Engineeringen
dc.date.accessioned2021-08-02T18:39:01Zen
dc.date.available2021-08-02T18:39:01Zen
dc.date.issued1987en
dc.description.abstractThe blood pumping chambers and drive pump for a totally implantable artificial heart have been developed at the biomedical engineering laboratory. The drive pump is driven by an electric motor and provides alternate pumping of the blood pumping chambers. The blood pumping chambers are diaphragm pumps and have uniquely captured valves. The pressure waveforms produced by the blood pumping chambers were measured at 26, 52 and 78 beats per minute and the hemolysis caused by the blood pumps was measured under pressures likely to be encountered in the body. The average hemolytic index over the three trials was 0.041 for the right blood pumping chamber and 0.040 for the left blood pumping chamber.en
dc.description.degreeM.S.en
dc.format.extentvii, 99 leavesen
dc.format.mimetypeapplication/pdfen
dc.identifier.urihttp://hdl.handle.net/10919/104538en
dc.language.isoenen
dc.publisherVirginia Polytechnic Institute and State Universityen
dc.relation.isformatofOCLC# 16679448en
dc.rightsIn Copyrighten
dc.rights.urihttp://rightsstatements.org/vocab/InC/1.0/en
dc.subject.lccLD5655.V855 1987.J37en
dc.subject.lcshElectric motorsen
dc.subject.lcshHeart, Artificialen
dc.titleThe design, development and in vitro testing of an electric motor driven artificial hearten
dc.typeThesisen
dc.type.dcmitypeTexten
thesis.degree.disciplineMechanical Engineeringen
thesis.degree.grantorVirginia Polytechnic Institute and State Universityen
thesis.degree.levelmastersen
thesis.degree.nameM.S.en